Radiation detected |
: |
|X and Gamma rays |
 |
Detector |
: |
A miniature halogen quenched G.M. Counter |
 |
Ranges |
:
: |
Exposure : 2 mSv full scale
Exposure rates : 0 - 200 mSv/hr
0 - 20 mSv/hr
0 - 2 mSv/hr
0 - 200 µSv/hr
0 - 20 µSv/hr |
 |
EHT |
: |
-550Volts. A flickering LED on the panel
indicates EHT ‘ON’. |
 |
Audio |
: |
Buzzer provides beeps for every pulse. Continuous
alarm when accumulated exposure
exceeds the preset level. |
 |
Energy response |
: |
Within +/-20% in the energy range 60 keV to 1.25
MeV. |
 |
Time constant |
: |
2.5 seconds on 20 µSv/hr range |
 |
Accuracy |
: |
Within 10% of reading between 10% and 100% of
full scale indication on any range. |
 |
Power requirement |
: |
6 Colts DC (4 x 1.5 Volts`AA’ type cells) |
 |
Weight |
: |
800 gms. |