OZONE MONITOR
![]() |
BARC has developed an ozone monitor for measuring ozone concentrations at ppb levels (10-9 v/v ratio) in the ambient atmosphere. It is based on chemiluminescent reaction of ozone with ethylene supplied from a cylinder. Light produced in this reaction in the region of 300-500 nm is measured by a sensitive photomultiplier tube and the signal is processed and displaced on the meter and recorded by a recorder or data processor. Light produced in the reaction is specific to ozone and other pollutants do not interfere.
SPECIFICATIONS
| Measuring ranges | : | 0-100 ppb |
| : | 0-1 ppm | |
| : | 0- 10 ppm | |
| Minimum detection limit | : | 5 ppb |
| Response time (10% to 90% rise) | : | 5 sec |
| Reproducibility | : | +1% |
| Linearity | : | +1% |
| Output | : | 0 - 1 volt (for air quality data processor) |
| Power requirements | : | 100 watt at 230 v, 50 Hz |
APPLICATIONS
Ozone monitor can be used for measuring ambient levels of ozone in polluted urban atmospheres and near high voltage power distribution centres.
For details contact :