Liquid scintillation counter
(LSC) is a high efficiency laboratory counting system. This system is used for the detection and quantification of radiations even with very low penetrating power such as
a
and b
particles. This is a PC based manually operated system for detecting even small amounts of beta and alpha radioactivity. Usage of excellent quality high efficiency matched pair of photo multiplier tubes, precision fast and slow preamplifiers, high speed analog electronics design, USB based spectroscopy quality MCA, rugged manually operated sample chamber with lead shield allow the system to offer high efficiency even for low energy beta emitters such as tritium and also a very low background.
SPECIFICATIONS
|
|
: |
Cocktail consisting of radioactivity dissolved in liquid
scintillator contained in glass or plastic vial viewed by matched pair of photo
multiplier tubes |
|
|
: |
Manually operated light tight sample chamber shielded
with lead |
|
|
: |
In 20 ml vials |
|
|
: |
Air conditioned
(20-250C) dust free Environment |
|
|
: |
System operates on 230V, 50 Hz, AC supply
High Voltage : Auto Set or User adjustable up to 1800V
Low Voltage : ±15V |
|
|
: |
Live display of acquisition of beta Spectrum |
|
|
: |
For tritium
(H3) (in toluene based scintillating solution) ³
50% with Background less than 20 cpm
For carbon (C14) (in toluene based scintillating solution)
³
90% with Background less than 35 cpm |
SALIENT FEATURES
- Advanced photo multiplier design
- Nanosecond pulses coincidence detection
- USB based high-resolution spectroscopy quality Multichannel Analyzer (1K -8K)
- Excellent performance in terms of efficiency, low background etc.
APPLICATIONS
- At Nuclear Power Plants to count
H3 activity of large number of samples
- At various environmental survey laboratories, hospitals and agricultural labs for measurement of radio labeled tritium and carbon
- Radio carbon dating
- Swipe or wipe assays
- General-purpose biomedical research using radiotracers
|