BAYARD-ALPERT GAUGES FOR UHV SYSTEMS

                                               Sealed Gauge                                                                              Nude Gauge

Introduction 

Bayard Alpert gauges (both sealed and nude types) are used for measurement of low pressure in the range of 10-4 to 10-9 torr. This is achieved by measuring the ionisation current, which decreases with the decrease in the pressure and is also dependent on the type of the gas species present in the vacuum system. These gauges come under the category of hot ionisation vacuum gauges. It consists of a filament, an ion collector and a grid. In sealed type gauges they are supported on a glass wafer base having a number of glass to metal (GM) seals. The metal pins apart from supporting the above components are also used to provide electrical connections. The whole assembly is housed in a glass envelope with wafer base fused at one end. In nude type gauges, a tubular Kovar metal to glass seal is welded to a standard SS flange. The tungsten filament, molybdenum grid and the tungsten ion collector are spot welded to the Kovar pins of the GM seal as in sealed type gauges. This assembly is protected by a removable transparent (acrylic) cover.  


Technical Specification

Filament : Tungsten
Filament voltage : 4.2 V
Filament current : 3.2 A
Filament potential : + 100 V DC
Collector : Tungsten
Collector potential : 0 V DC
Grid and Supports :  Molybdenum
Grid potential : + 200 V DC
Degass power for grid : 30 to 40 Watts
Pressure range : 10-4 to10-9 torr
Sensitivity : ~ 12/ torr for Nitrogen
Leakage current : ~ 10-12 A
Envelope and tubulation material (For Sealed Gauge) :  Kodial sealing glass
Baking temperature: nude gauge/ sealed gauge :

4000 C/ 3500 C

Flange size for nude gauge : CF 35 (70 mm OD Conflat)
Flange material : SS 304
Protection (For Nude gauge) : Perspex cover

 

Application 

These gauges are required for the measurement of vacuum in the range 10-4 to10-9 torr. Such gauges are needed for mass spectrometers, surface analytical instruments, thin film deposition units, accelerators and all other equipments/ processes needing ultra high vacuum. Apart from DAE, research institutions/ universities and some of industrial units using UHV systems may find these vacuum gauges very useful.