Cardiac Output Monitor reveals pump performance of the heart. The board provides Cardiac Output signal in analog form which can be displayed on standard Oscilloscope. It will form front end module to Patient Monitoring systems useful for the monitoring of stroke volume and cardiac output in patients. It is based on the principle of electrical impedance plethysmography and is non-invasive in nature.
This method has been reported to be accurate in majority of subjects except those with intracardiac shunts or valvular regurgitation and at present is the only non-invasive method for continuous cardiac output monitoring.
SPECIFICATIONS
| Principle |
: |
Electrical impedance measurement by constant current
method. |
| Mode of operation |
: |
Calibration/Patient |
| Configuration |
: |
Tetrapolar |
| Sensor |
: |
Spot electrodes |
| Carrier current |
: |
Sinusoidal at 48 KHz frequency and 4 mA amplitude |
| Measurement range |
: |
Impedance 5 to 75 W |
| Rate of change of impedance |
: |
-1.5 to +3.5 W/sec |
| Stroke volume |
: |
10 ml to 150 ml |
| Cardiac output |
: |
0.5 litres to 20 litres |
| Patient isolation |
: |
3000 Volts |
| Power input |
: |
230 VAC +/- 10%, 50 Hz, 100 W |
| Output |
: |
On Oscilloscope |
APPLICATIONS
This can be used for OEM application for continuous monitoring of stroke volume and cardiac output in critically ill patients in Intensive Care Units and Intensive Cardiac Care Units, with hardware addition.
|